Efficacy and safety of proact for the treatment of male post-surgical stress urinary incontinence: a systematic review of literature

Orecchia L1, Iacovelli V1, Farullo G2, Turbanti A2, Petta F1, Vespasiani G1, Finazzi Agrò E1

Research Type

Clinical

Abstract Category

Male Lower Urinary Tract Symptoms (LUTS) / Voiding Dysfunction

Abstract 72
Open Discussion ePosters
Scientific Open Discussion Session 7
Wednesday 29th August 2018
12:30 - 12:35 (ePoster Station 1)
Exhibition Hall
Male Stress Urinary Incontinence Surgery
1. University of Rome Tor Vergata, Department of Experimental Medicine and Surgery, Urology Unit, Rome, Italy, 2. Urology Residency Program, University of Rome Tor Vergata, Rome, Italy
Presenter
Links

Poster

Abstract

Hypothesis / aims of study
Stress Urinary Incontinence (SUI) is one of the most clinically significant complications following prostate surgery, with rates varying between 3%-74% for post-prostatectomy incontinence and 0%-0.5% for post-TURP incontinence. When conservative management fails to re-establish continence, the recommended treatment is implantation of a urinary incontinence device. The aim of this study is to perform a systematic review of the evidence regarding the male Adjustable Continence Therapy (ProACT) device, with focus on its functional outcomes and complications in the treatment of SUI.
Study design, materials and methods
A computerised systematic search of papers published from 1998 in English language was independently conducted by two of the authors on three databases [MEDLINE (Via Pubmed), Scopus and ISI Web of Science] in January 2018. Search terms included: “Adjustable Continence Therapy”, “Adjustable Continence Balloons”, “ProACT” and “Periurethral Balloons”. 814 records were identified through the database search. Abstract-only publications, conference papers and reviews were deemed not eligible. To be eligible, studies had to have minimum follow-up of 12 months, cohort size of more than 20 consecutive patients, complete reporting of both outcomes and complications with the device, and no other urinary incontinence device implanted at the time of ProACT implantation. After removing duplicate results and assessing eligibility based on outlined requirements, 11 articles were included in the qualitative synthesis (Fig.1). Systematic review and data extraction were conducted independently and then cross-checked by two authors using data extraction forms. No disagreement during the inclusion process occurred.
Results
(Fig.2) Of the 11 studies selected for review, 9 were prospective single centre studies, 1 was a prospective multicentre study, and 1 was a retrospective single centre study. These studies involved 833 patients in total, with a mean/median time of follow-up up to 58 months. Most patients had the ProACT device implanted to treat post-prostatectomy SUI (59%-100%). In all studies, the primary outcome assessed was reduction in pad per day usage. Most of the studies also assessed changes in the I-QoL score after implantation. All studies reported perioperative and long-term complications with the device.
Success rates varied between 45%-71% (definition of success: 0 pad or 1 pad per day postoperatively, 8 studies) and 4.5%-68% (definition of success: 0 pad per day postoperatively, 3 studies). The mean number of pads per day ranged from 2.8 to 5.9 preoperatively and 0.75 to 3.9 postoperatively (10 studies). The mean I-QoL score ranged from 31.7 to 61.0 preoperatively and 66.3 to 84.3 postoperatively (10 studies). The main perioperative complications recorded were: urethral perforation (up to 11.9%), bladder neck perforation (up to 9%), and acute urinary retention (up to 5%), No study reported significant perioperative bleeding. The main postoperative complications recorded were: migration/dislocation of the device (up to 14%), urethral erosion (up to 11.1%), infection (up to 8%) and balloon rupture/loss of volume (up to 15.4%). Overall the rate of revisional surgical procedures ranged from 6.3% to 34.3% (9 studies). Complications and failure rates were higher in sub-groups of patients who received radiotherapy after the prostatectomy. Postoperative complications were ranked according to the Clavien-Dindo Classification in 2 studies: one reported 3 (2.1%) grade IIIb complications in the first 30 postoperative days, the other reported 29 grade IIIb complications in 22 patients. In 3 studies, reported data allowed to assess that no grade IIIb or higher complication occured, whereas grade IIIa complications occurred in 7.1% to 20.2% of the patients. None of the included studies reported complications rankable as grade IV or higher. The reported mean operative time ranged from to 19 to 69 minutes (7 studies).
Interpretation of results
The studies selected include mixed patient populations in terms of ethiology of the SUI, severity of the incontinence and history of adjuvant radiotherapy. Howbeit the functional outcomes of ProACT and AUS are similar and seem to be superior to the ones of the male slings. Even though the AUS remains the gold-standard treatment, its complication rates are higher when compared to ProACT. Randomised trials should be conducted to compare the different SUI treatments in terms of efficacy, long-term safety and durability.
Concluding message
ProACT has low complication rates and high success rates when compared to other urinary incontinence devices. Thanks to its adjustability, minimally invasive design and short operative times, it may have the potential to be the first-line treatment of choice for males with any degree of post-surgical SUI.
Figure 1
Figure 2
Disclosures
Funding None Clinical Trial No Subjects Human Ethics not Req'd The study is a systematic review of literature. Helsinki Yes Informed Consent Yes