The BLAI volume measurement was found to be non-inferior (p<.001), with a mean difference between percent differences of the human measurement and BLAI means of -0.0228 (95% CI -0.074, 0.028). Bland-Altman plots were assessed to determine agreement between the two methods of bladder volume measurement. The mean difference and tight clustering of bladder volume measurements between BLAI and human measurements indicated strong overall agreement with individual human measurements. The ICC showed strong agreement between reviewer 1 vs reviewer 2 of 0.98 ( 95% CI 0.96, 0.99), reviewer 1 vs reviewer 3 of 0.98 (95% CI 0.97,0.99), reviewer 2 vs. reviewer 3 of 0.98 (95% CI 0.97,0.99) (Table 1). Additionally, the ICC comparing BLAI to the reviewer mean was 0.91 (95% CI 0.85,0.95) (Table 1). The average Dice coefficient and Jaccard index scores showed strong agreement between the BLAI model compared to each reviewer and between each reviewer pair. Dice scores for reviewers 1, 2, and 3 compared to BLAI were 0.93 (95% CI 0.92, 0.93), 0.92 (95% CI 0.91, 0.93), and 0.93 (95% CI 0.92, 0.93), respectively. Comparison pair Dice scores for reviewer 1 vs. reviewer 2 was 0.93 (95% CI 0.91, 0.94), reviewer 1 vs reviewer 3 was 0.94 (95% CI 0.93, 0.94), and reviewer 2 vs reviewer 3 was 0.94 (95% CI 0.93, 0.95). Jaccard scores for reviewers 1, 2, and 3 compared to BLAI were 0.87 (95% CI 0.85, 0.88), 0.86 (95% CI 0.84, 0.87), and 0.86 (95% CI 0.85, 0.88) respectively. Comparison pair Jaccard scores for reviewer 1 vs. reviewer 2 was 0.87 (95% CI 0.85, 0.88), reviewer 1 vs reviewer 3 was 0.89 (95% CI 0.87, 0.90), and reviewer 2 vs reviewer 3 was 0.88 (95% CI 0.87, 0.90)(Table 2). A sub-group analysis compared bladder volumes measured with Clarius C3HD with an ICC of 0.90 (95% CI 0.81, 0.94) and Phased array with an ICC of 0.97 (95% CI 0.92, 0.99), thus demonstrating excellent reliability.